Xtester- HT-PA310 Portable Oxygen Analyzer
Technical Parameter:
Measuring principle: | Zirconia |
Display mode: | 128×64 dot matrix LCD |
Measuring range: | 0~10/ 100/ 1000PPm/ 1.00%/25.00% O2 |
Measurement accuracy: | 0~1000PPm/ 1.00%/25.00%≤±1%FS 0~100PPm≤±2%FS 0~10PPm≤±5%FS |
Resolution: | 0.1PPm (trace) 0.01% (constant) |
Repeatability: | ≤±1.0%FS |
Response time: | T90≤30S |
Ambient temperature: | -10℃~+50℃ |
Storage environment humidity: | <90%RH, non-condensing |
Working environment humidity: | <100%RH, non-condensing |
Sample gas temperature: | 0℃~60℃ |
Sampling method: * | inlet type or suction type (please specify when ordering) |
Sample gas flow: | 400~600mL/min |
Inlet pressure: | Inlet type: 5kPa≤relative pressure≤200kPa (>200kPa is recommended to install a pressure reducing valve) Air extraction type: -10kPa≤relative pressure≤5kPa (when <5kPa, please specify the specific negative pressure value when ordering, determine its gas path structure) |
Discharge pressure: | free emptying (under safe conditions) |
Background gas: | N2 or other inert gas |
Charging power supply: | AC100~240V 50/60Hz to DC12V 2.5A Battery capacity: 14.8V,11AH |
Size: | 155mm×260mm×340mm (H×W×D) |
Sensor life: | > 5 years (under normal use conditions) |
Pneumatic interface: | NPT 1/8 internal thread |
Installation method: | portable |
Instrument features
1. Large-screen dot-matrix liquid crystal display, which can simultaneously display parameters such as oxygen level, date, time, etc.; |
2. Adopt new long-life zirconia sensor, which has the characteristics of no power consumption, no consumption of life, high sensitivity, fast response speed and high precision; |
3. It can work in the air for a long time without affecting the life of the sensor; |
4. The response speed is fast, and the minimum measurement is 0.1PPm; |
5. Automatic temperature compensation function to eliminate the influence of sample gas temperature change on measurement accuracy; |
6. 5000 pieces of historical data are automatically stored, allowing customers to freely check historical data; |
7. Powered by a large-capacity lithium battery, the continuous measurement time is more than 24 hours, and the standby time is more than 168 hours; |
8. The instrument supports multi-point segmented linear calibration, which can meet the accurate measurement of oxygen content in the full range; |
9. Built-in original imported sampling pump (optional), long life and reliable work. |
Applications
Widely used in the detection and analysis of oxygen content in air separation nitrogen production, smelting industry, medical and health, petrochemical, electronic power and other industries. |