Contact Us For Quotation | Email: 464560351@qq.com | Whatsapp&Wechat:+8618620373879 |
Xtester-VCT160N 1.0x10-2~1.0x10+5 Pa Pirani Vacuum Gauge
Introduction
VCT160N is a highly cost-effective product developed specifically for measuring medium and low vacuum environments The structure design of the Farani vacuum gauge, which separates the probe from the instrument, enables the use of VCT160N to be compatible with the instrument The maintenance cost is much lower than that of similar vacuum gauges. Unique control instruments are particularly suitable for industrial control systems The cabinet installation mode in can also be placed separately on the platform for observation and reading. Compared to traditional vacuum gauges, VCT160N not only has a more aesthetically pleasing exterior design, but also has needles Multiple functional optimizations have been carried out for sex, resulting in excellent reliability, ease of use, and convenience and economy. | |||||
Features
1,Separate design: The instrument and probe can be replaced independently |
2,Vacuum degree range covers 1E-2Pa ⾄ 1.0E+5Pa |
3,Intelligent reminder of vacuum gauge operation status and health status |
4,Intelligent Calibration: Automatic Recognition of Vacuum Zero Position |
5,Full range temperature compensation: Vacuum degree measurement is more stable and accurate |
6,Multiple units available: Pa, mbar, torr can be switched |
7,Multi display mode: conventional counting method and Scientific notation can be switched |
8,Strong control function: optional four single/double throw relay outputs, which can achieve vacuum degree Single point and interval control mode. |
9,Dual communication mode, standard data communication (Modbus), optional analog output (0-10V and 4-20mA) |
10,Data storage and analysis: compatible with IVMS vacuum degree data recording and analysis system Collaborate to achieve real-time data collection, recording, and analysis |
11,Support the fault manual release function and firmware upgrade function |
Specification
Materials exposed to vacuum | SS316L.Covartungsten,glass,Stainless steel, tungsten, kovar |
Measurements Range | 1.0x10-1~1.0x10+5 Pa(Air, N2, adjustment is necessary for other gas) |
Accuracy(% of reading) | 1.0x10-2~1.0x10-1Pa:+30% 1.0x10-1~10x10+4Pa:+10% 1.0x10+4~10x10+5Pa:+30% Equal: 1.0x10-4~1.0x10-3mbar:+30% 1.0x10-3~10x10+2mbar:+10% 1.0x10+2~10x10+3mbar:+30% |
Display accuracy(% of reading) | +/-10%; zero shifts: B29+/-5% |
Data acquisition | Digit resolution: 1% ; response time: <100ms; update rate of display: 1s |
Inputs | Four press buttons for: units selection、calibration for Atm and high vacuum.setpoints |
Output | Voltage/Current analog, RS485Modbus RTU |
Control unit | Four way SNDT relay; load: 3A/220VAC,non inductive load; response time:<1s ;Setpoint remembering when power lose |
Admissible temperature | Operation:0C~+45°C;Storage:-40C~+75°C Vacuum connection parts: 180C Max (electronic unit removed) |
Power consumption | 10W Max |
Power supply | +85VDC~265VAC(20%)/0.5Ais recommended |
Electrical connection | DB Sub9, 9-pin malelanalog output, RS485 output, power supply, identification) |
Vacuum connection | DN 16 ISO-KF(other optional) |
Weight | 0.5KG (including 2 DN 16ISO-KF and 2 cable with 3m length) |
Mounting type | Embed hole: 91X91(+0.2/-0.0)mm |