Xtester-Fluke 753 Documenting Process Calibrator

Xtester-Fluke 753 Documenting Process Calibrator

2023-07-19 15:45:58  

Xtester-Fluke 753 Documenting Process Calibrator

  • Main Features : Full-featured documenting process calibrator for calibrating and troubleshooting Measures volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments Sources and simulate volts, mA, pressure, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters Powers transmitters during test using loop supply with simultaneous mA measurement
  • Serial No. :
Contact Us for QuotationEmail:464560351@qq.comWhatsapp: +8618620373879

Product Introduction

  The Fluke 753 is a powerful multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software, and upload data for printing, archiving, and analysis. You'll find it does the work of several tools, with simultaneous source and measurement capabilities for all common process parameters. It sources, simulates, and measures pressure, temperature, and electrical signals in one rugged, hand-held calibration tool. It also automates calibration procedures, captures data for documentation, and helps you meet rigorous standards like ISO 9000, FDA, EPA, and OSHA regulations.

  The 753 includes a USB interface and a USB communication cable to enable two-way communication with DPCTrack 2 and other instrumentation management applications. To create a seamless/paperless calibration management system consider adding Fluke DPCTrack2 Calibration Management software for use with your Fluke 753 and 754 or even legacy Fluke 743 and 744 calibrators.

Specification

Measurement Accuracy
Voltage DCRange /Resolution1 Year2 Years
100.000 mV0.02% + 0.005 mV0.03% + 0.005 mV
3.00000 V0.02% + 0.00005 V0.03% + 0.00005 V
30.0000 V0.02% + 0.0005 V0.03% + 0.0005 V
300.00 V0.05% + 0.05 V0.07% + 0.05 V
Voltage AC3.000 V (40 Hz to 500 Hz) / 0.001 V0.5% + 0.002 V1.0% + 0.004 V
30.00 V (40 Hz to 500 Hz) / 0.01 V0.5% + 0.02 V1.0% + 0.04 V
300.0 V (40 Hz to 500 Hz) / 0.1 V0.5% + 0.2 V1.0% + 0.2 V
Current DC30.000 mA0.01% + 5 uA0.015% + 7 uA
110.00 mA0.01% + 20 uA0.015% + 30 uA
Resistance10.000 Ω0.05% + 50 mΩ0.07% + 70 mΩ
100.00 Ω0.05% + 50 mΩ0.07% + 70 mΩ
1.0000 kΩ0.05% + 500 mΩ0.07% + 0.5 Ω
10.000 kΩ0.1% + 10 Ω0.15% + 15 Ω
Frequency1.00 to 110.00 Hz / 0.01 Hz
0.05 Hz
110.1 to 1100.0 Hz / 0.1 Hz
0.5 Hz
1.101 to 11.000 kHz / 0.001 kHz
0.005 kHz
11.01 to 50.00 kHz / 0.01 kHz
0.05 kHz
Source Accuracy


1 Year2 Years
Voltage DC100.000 mV0.01% + 0.005 mV0.015% + 0.005 mV
1.00000 V0.01% + 0.00005 V0.015% + 0.0005 V
15.0000 V0.01% + 0.0005 V0.015% + 0.0005 V
Current DC22.000 mA (source)0.01% + 0.003 mA0.02% + 0.003 mA
Current sink (simulate)0.02% + 0.007 mA0.04% + 0.007 mA
Resistance10.000 Ω0.01% + 10 mΩ0.015% + 15 mΩ
100.00 Ω0.01% + 20 mΩ0.015% + 30 mΩ
1.0000 kΩ0.02% + 0.2 Ω0.03% + 0.3 Ω
10.000 kΩ0.02% + 3 Ω0.03% + 5 Ω
Frequency0.1 to 10.99 Hz
0.01 Hz
0.01 to 10.99 Hz
0.01 Hz
11.00 to 109.99 Hz
0.1 Hz
110.0 to 1099.9 Hz
0.1 Hz
1.100 to 21.999 kHz
0.002 kHz
22.000 to 50.000 kHz
0.005 kHz
Technical Data
Data log functionsMeasure functionsVoltage, current, resistance, frequency, temperature, pressure
Reading rate1, 2, 5, 10, 20, 30, or 60 readings/minute
Maximum record length8000 readings (7980 for 30 or 60 readings/minute)
Ramp functionsSource functionsVoltage, current, resistance, frequency, temperature
Rate4 steps/second
Trip detectContinuity or voltage (continuity detection not available when sourcing current)
Loop power functionVoltageSelectable, 26 V
Accuracy10%, 18 V minimum at 22 mA
Maximum current25 mA, short circuit protected
Maximum input voltage50 V DC
Step functionsSource functionsVoltage, current, resistance, frequency, temperature
Manual stepSelectable step, change with arrow buttons
AutostepFully programmable for function, start delay, stepvalue, time per step, repeat
Environmental Specifications
Operating temperature-10°C to +50°C
Storage temperature-20°C to +60°C
Dust/water resistanceMeets IP52, IEC 529
Operating altitude3000 m above mean sea level (9842 ft)
Safety Specifications
Agency approvalsCAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993
Mechanical and General Specifications
Size136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in)
Weight1.2 kg (2.7 lb)
BatteriesInternal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh
Battery life> 8 hours typical
Battery replacementVia snap-shut door without opening calibrator; no tools required
Side port connectionsPressure module connector
USB Connector to interface to your PC
Connection for optional battery charger/eliminator
Data storage capacity1 week of calibration procedures results
90 day specificationsThe standard specification interval for the 750 Series are 1 and 2 years.
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2.
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant.
Temperature, Resistance Temperature Detectors
Degrees or % of reading - Type (α)Range °CMeasure °C1

1 year2 year
100 Ω Pt (385)-200 to 100
100 to 800
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
200 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
500 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
1000 Ω Pt (385)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3916)-200 to 100
100 to 630
0.07°C
0.02% + 0.05°C
0.14°C
0.04% + 0.10°C
100 Ω Pt (3926)-200 to 100
100 to 630
0.08°C
0.02% + 0.06°C
0.16°C
0.04% + 0.12°C
10 Ω Cu (427)-100 to 2600.2°C0.4°C
120 Ω Ni (672)-80 to 2600.1°C0.2°C
Source currentSource °CAllowable current2

1 year2 year
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
500 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
250 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
150 μA0.06°C
0.017% + 0.05°C
0.12°C
0.034% + 0.10°C
0.1 mA to 1 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
1 mA0.05°C
0.0125% + 0.04°C
0.10°C
0.025% + 0.08°C
0.1 mA to 10 mA
3 mA0.2°C0.4°C0.1 mA to 10 mA
1 mA0.04°C0.08°C0.1 mA to 10 mA
1. For two and three-wire RTD measurements, add 0.4°C to the specifications.
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms