Contact Us For Quotation | Email: 464560351@qq.com | Whatsapp&Wechat:+8618620373879 |
Xtester-Rotronic HC2A-IC302 Temperature And Humidity Probe
Characteristics of HC2A-IC302
• Scope of work:- 100…200 °C / 0…100 %rh
• Digital interface (UART) and variable range analog output, 0…1 V
• Standard output range: 0…1 V = -40…60 °C / 0…100 %rh
Calibrate 10 at 23 ° C, 35, 80% RH is suitable for heavy industry applications installed through walls
ROTRONIC industrial probes are particularly suitable for high temperature and harsh industrial environments. Measure humidity and temperature, and calculate dew/frost points.
Application
Production environment, high temperature environment, industrial manufacturing, drying process, climate chamber
characteristic
• Scope of work:- 100…200 °C / 0…100 %rh
• Digital interface (UART) and variable range analog output, 0…1 V
• Standard output range: 0…1 V = -40…60 °C / 0…100 %rh
Calibrate 10 at 23 ° C, 35, 80 %rh
Accuracy ± 0.8% RH/± 0.1 ° C at 23 ± 5 ° C
Accuracy, standard accuracy mode at 23 ° C/10, 35, At 80% RH, ± 0.8% RH/± 0.1 ° C probe type industrial probe
The scope of work is -100... 200. C / 0…100 %rh
Measurement range -100... 200. C (short-term peak load) / 0...100 %rh
Power supply/power consumption 3.3 V ± 0.1V/~4.5 mA Long term stability<1% rh/year
Sensor Hygomer IN-1
Temperature sensor PT100 1/3 DIN Class B probe protection filter bracket, brass nickel plated
Filter cartridges can be ordered separately based on the application, with a response time of T 63<15 seconds
Maximum wind speed of 3 m/s, calculated dew point or frost point
The repeatability of AIRCHIP is 0.02% RH/0.01. C Development Software ROTRONIC HW4
Adjust probe 1 point&multipoint% rh&° C firmware via software Upgrade via Internet and HW4 software
User information comes from HW4 (Device Management)
Data processing with interface cable via HW4
The device is password protected
User defined analog output has
Analog output signal (standard) 0... 1 V=0... 100% rh; 0... 1 V=-40... 60. C
Digital output signal (standard) UART
SENSOR FRAME FOR %RH AND °C SENSOR Yes
Output signal type 2 x 0... 1 VDC
The maximum cable length is 5m for passive cables and 100m for active cables. The probe material is PEEK
Reverse voltage protection mechanical connection service interface UART
FDA/GAMP compatibility compatible with FDA 21 CFR Part 11 and GAMP size Ø 15 x 250 mm
Weight 260 g |