China Testing Instruments Manufacturer,Whatsapp/Wechat:+8618620373879  Email: 464560351@qq.com

Xtester-PT720 (100kHz,2000A)AC Current probe

Xtester-PT720 (100kHz,2000A)AC Current probe

2023-07-13 13:47:15  

AC Current probe Bandwidth:100kHz 3 ratio:100mV/A,10mV/A,1mV/A 3 current range:0.1A-10A, 10A-200A ,200A-2000A

  • Main Features :
  • Serial No. :
Contact Us for QuotationEmail: mewoi008@mewoi.comWhatsapp:+8619928420370

Xtester-PT720 (100kHz,2000A)AC Current probe


INTRODUCTION

2000A high precision linear measurement, evenly distributed coils minimize phase shift, oscillating magnetic element adjustment system, can clamp 52mm conductors and micrometal coils, suitable for power meters. With insulated BNC coaxial cable, with any multimeter, meter recorder, oscilloscope, etc., easy to display and measure the current waveform, output mV signal. It has 3 ranges.

Technical Specification

Current range

0.1A~10A

10A~200A

200A~2000A

Attenuation ratio

100mV/A (10:1)

10mV/A (100:1)

1mV/A (100:1)

Accuracy

≤3%+10mV

≤2%+5mV

≤1%+1mV

Bandwidth

10Hz~100kHz(-3dB)

Max. dV/dt

10A/μs

Max. current

when >1kHz,1000A   Continuous current;If ≥1200A(no longer than 20 minutes)

Load impedance

≥1MΩ and ≤47pF

Rise/Fall time

≤3.5us

Working voltage

600V rms

Common mode voltage

600V CAT III

Stack DC current influence

≤30A DC

Max. Jaw opening diameter

53mm

Max. Conductor size

Cable:max. Diameter:Ф52mm;Total wiring harness;single:5х5mm,Four bus wiring harness:30х5mm

Size

216х111х45mm

Weight

500g

Selection Guide

ModelXtester-PT-720Xtester-PT-721
Current range0.1A-10A  |  10A-200A  |  200A-2000A0.1A-10A  |  10A-200A  |  200A-2000A
Output conversion ratio100mV/A(10:1) | 10mV/A(100:1) | 1mV/A(1000:1)100mV/A(10:1) | 10mV/A(100:1) | 1mV/A(1000:1)
Electrical indicators
The output error≤3%+10mV  |  ≤2%mV+5mV  |  ≤1%+1mV≤3%+10mV  |  ≤2%mV+5mV  |  ≤1%+1mV
Measurement conditions23C±3C, 20~75%RH, sinusoidal frequency 48Hz~65Hz, distortion degree <1%, no DC component, external magnetic field <40A/m, no AC magnetic field, central test sample, load impedance ≥1MΩ and ≤47pF
Bandwidth10Hz-100kHz(-3dB)
Max. dV/dt10A/ us
Maximum currentAt frequencies >1kHz, 1000A continuous current (beyond this frequency limit ratio): 1200A approximately 20 min maximum test
The load impedance≥1MΩand≤47pF
Rise/Fall time≤3.5us
Working voltage600V rms
The common mode voltage600V CAT III and interference level 2
Effect ofsuperimposed DC current≤30A DC, < output signal 1%
Diameter ofmaxillary mouth53mm
Maximumconductor sizeCable: maximum diameter 52mm; Bus harness; Single river flow bar 5x 5mm, four bus bar 30 x 5mm
Tong head size216x111x45mm
Temperature andhumidity0-50C / 20-75% RH


Telephone
Email
Map