Contact Us For Quotation | Email: 464560351@qq.com | Whatsapp&Wechat:+8618620373879 |
Xtester-VM-E50 Drying Method Halogen Moisture Analyzer
Introduction
VM-E series "drying method" halogen moisture analyzer using high efficiency drying heater - high quality cyclic halogen light, the lamp heats the sample quickly and evenly, and the moisture of the sample is continuously dried. The entire measurement process is fast, automatic and simple. The instrument displays the measurement results in real time: moisture value, solid content, water weight dry weight ratio, wet weight dry weight ratio, sample initial value, final value, measurement time, the final temperature value, trend curve,etc. The instrument can input data to printer, computer, PLC, etc. through RS232 serial port, which can be used in a few minutes (most samples), quickly measure the moisture in solid, liquid and powder, easy to operate, integrate multiple heating modes, and work in a variety of ways. Demand is widely and used in industries such as Grain, food, agriculture, chemicals, plastics, new energy, new materials, medicine, scientific research and testing etc. |
VM-E Series " drying method " halogen moisture analyzer principle:
The heating weight loss method is a kind of physical analysis method for moisture determination, which uses the influence of temperature on water volatilization to dry out the moisture in the material. The internationally accepted standard measurement method is the oven method. As the name suggests, the material sample is placed in an oven for drying and heating, and then weighed by an electronic balance. The moisture value is calculated manually. VM-E10 Halogen Moisture Analyzer Series distinguish it from other products is that the traditional moisture measurement is used oven drying method. The test of one sample takes two to three or even three to four hours, and it needs to be weighed manually and calculated by the balance to get the moisture value of the sample. The low efficiency of the traditional oven method for moisture determination cannot adapt to the high - tempo enterprise production needs. The moisture meter use halogen light as heating source, so called Halogen Moisture Analyzer, VM- E10 principle is implantation halogen gas such as iodine or bromine in the bulb, at high temperature effect. The sublimated tungsten filament chemically interacts with the halogen, the sublimated tungsten will re- solidify on the tungsten wire to form a balanced cycle to prevent premature rupture of the tungsten wire. Therefore, halogen bulbs are longer than incandescent lamps. At the same time, because the filament can work at higher temperatures, so higher brightness, higher color temperature and higher luminous efficiency are obtained. VM-E10 " drying method " Halogen Moisture Analyzer integrated upgrade the principle of international oven drying method, the heating module, the weighing module, the operation module and the human-machine intelligent module are perfectly combined, measurement trend and the measurement result system are automatically determined and automatically stopped. The results are in good agreement with the oven method for moisture determination, and the work efficiency is much higher than the oven method for moisture determination. Use VM- E10 " Drying" method halogen moisture meter, the general sample can be measured in just a few minutes, so it is favored and praised by enterprise users. |
Applicable Industry
General industry | food, wood, paper, textile, coal, ore, sand, soil, fertilizer, feed, chemical, tobacco |
Inorganic salt | calcium carbonate, refractory material, talcum powder, alumina, gypsum, desiccant, eramic slurry, ceramic powder |
Food and Drug | Chinese medicine, western medicine, health food, essential oil, dehydrated food, food stuffing, food powder, fruits and vegetables, dried fruit, noodles, sugar, meat, cooking oil, etc. |
Battery industry |
lithium battery powder, Lithium battery paste, battery class, graphene. |
Plastic rubber industry | ABS, PP, PVB, PC, COC, BMC, DMC, SMC, BOPP, PBT plastic film, ESP polylon, PET plastic, PA polyamide, LCP, PVC, recycled plastic, bottle flakes, PE polyethylene, rubber, PS styrene, EPE pearl wool |
Solid content | paint, milk and soy milk, glue, adhesive, rubber, slurry, mud, sludge, silica gel, chocolate, soy sauce, toothpaste, etc. |
Our company has a series of moisture measurement technology solutions, welcome to inquire your exclusive moisture measurement solution! |
Technical Parameters
Model | VM-E100 | VM-E50 | VM-E20 | VM-E10 | VM-E01 |
Weighing Accuracy (d) | 0.01g | 0.005g | 0.002g | 0.001g | 0.0001g |
Moisture Readability | 0.01% | 0.001% | |||
Moisture Range MC% Solid Content Range DC% | 0.00% - 100.00% | 0.000% - 100.000% | |||
Weight Sensor | HBM Strain Gauge Force | Electromagnetic Force | |||
Weighing Range | 128g (optional 224g/300g) | ||||
Calibration Method | External Weight Calibration (Single-Point & Multi-Points) | ||||
Repeatability (Standard Deviation) | ≤3d | ||||
Heating Source | Halogen Lamp | ||||
Temperature Sensor | PT- 1000 Temperature Transmitter | ||||
Heating Program | Standard Heating, Step heating, Rapid heating, Soft Heating | ||||
Heating Temperature Range | 40-250℃ | ||||
Temperature Accuracy | 0.1℃ | ||||
Heating Time setting | 1-99 minutes by 1 minute | ||||
Stop Mode | Timing, Manual, Automatic | ||||
Measurement Data | 2004 sets | ||||
Display Data | MC%(Moisture),DC%(Solid Content), ATRO Dry, ATRO Wet,Sample Weight,Moisture Weitht,Dryer Sample Weitht,Tempreture,Time, Analysis Curve | ||||
Function | Custom Measurement Method, U Disk System Upgrade | ||||
Display& Operating | 5-Inch Capacitive Touch Screen | ||||
Sample Tray | 50 aluminum sample trays with 90mm diameter, optional stainless steel sample tray | ||||
Sample Height | 24.5mm | ||||
Sample Pan | Aluminum Sample Pan(50pcs), Stainless Steel Sample Pan(4pcs) | ||||
Thermal Protection | Aluminum High Temperature Radiator | ||||
Baud rate | 2400,4800,9600,19200 | ||||
Data Communication | RS232/RS485(can be connected to a computer, printer, PLC, etc.) | ||||
Print Function | Optional Printer (thermal, needle, adhesive, etc.) | ||||
Use Environment | 1-38℃ (18-25℃ ideal temperature) | ||||
Power | Voltage 220v±10%, frequency 50HZ±1HZ (110V power supply and power cord can be customized for export) | ||||
Host Size (W*D*H) | 315mm*210mm*205mm | ||||
Voltage | 220V or 110V 50HZ Max 500W | ||||
Weight | 5.8KG | 8.8KG |